2011-0816 Premier Orthopedic TI.pdf City of Edmonds PLAN REVIEW COMMENTS BUILDING DIVISION (425) 771-0220 DATE: November 7, 2011 TO: Scott Shanks First Western Development Services, Inc. Email: Scott@fwdsinc.com FROM: Leif Bjorback, Assistant Building Official RE: Plan Check: 2011-0816 Project: Premier Orthopedic TI nd Project Address: 21401 72 Ave W During review of the plans for the above noted project, it was found that the following information, clarifications or changes are needed. Please provide written responses as to where the changes can be found on the plans, and submit revised plans/documents to Marie Harrison, Senior Permit Coordinator. Thank you. 1.Please provide a copy of the Washington State DOH plan review letter. 2.Provide fire stopping specifications and details for through penetrations of fire rated assemblies greater than one hour, or list as a deferred submittal. 3.The north wall of the elevator hoistway looks to be considered the boundary of the atrium. The opening through this wall at the third floor must be protected by a 45 minute fire rated door. IBC 404.6 and Table 715.4 4.The window at MRI Room 375 does not match the plans for the building shell. This window must be 10 feet away from the adjacent atrium windows. IBC 1022.6 5.Please indicate on the plans that all fire doors will be self closing or automatic closing. IBC 715.4.8 6.Please indicate that all fire doors will be smoke gasketed. IBC 715.4.3.1 7.The door to the med gas room must be rated for at least 45 minutes. IBC Table 715.4 8.Indicate that the med gas system will be installed according to UPC chapter 13 and inspected by a third party as per UPC section 1328. 9.Provide trap primers at restrooms with floor drains. UPC 1007.0 10.It appears the outside air calculated for Waiting Room 306 is insufficient. Using the square footage of 950 provided on sheet M0.01, and IMC Equation 4-1, my calculations show a minimum requirement of 1911 cfm for that room. Please show new calculations for the room. 11.Provide a smoke damper at the air transfer opening through the smoke partition at the level 2 lobby. See sheet M2.02. IBC 711.7 nd 12.Provide emergency lighting on the 2 floor level. See sheet E3.2. IBC 1006.3 rd 13.Clarify the emergency lighting on the 3 floor level. The symbols used on sheet E3.3 don’t seem to match those in the key. IBC 1006.3