Gas Outlet Report.pdfG (- W 6,0 L b rz t~L
Medical Gas Services, LLC
Medical Gas Outlet / Inlets Evaluation Worksheet
Facility: Swedish MC - Edmonds Job #: I Date: 12/8/17
Area Tested: PCU East Tower Verifier: HP
7nna Valva. n Nnna n Nnf Rpnnirarl n Naw n Fxi0ina
Labeled As: PCU Room 201
ZV Brand: Chemetron
Gas
3 Part Valve
DL Gauge
Ga. PSI
Act PSI
Comments:
OX
❑ Yes ❑ No
❑ Yes ❑ No
MA
❑ Yes ❑ No
❑ Yes ❑ No
VAC
❑ Yes ❑ No
❑ Yes ❑ No
N20
® Yes ❑ No
® Yes ❑ No
58
56
Aran Alarm - F— Mnnn I-1 MM Rarmirorl n Maw mC FYictinn
Labeled As: No label
Brand: Chemetron
Gas
PSI
Set HI
Set LO
Sensor
Comments:
OX
MA
VAC
N20
54.5
60
40
UL
nidinfe R Inlefe . F-1 Maw r7l Fvictinn
Outlet / Inlets Brand: Chemetron
Location: Wall
Quick Connect: Chemetron
Area / Room / Door
Gas
Flow
SCFM
Delta
Flow
Conc.
%
Part.
Comments:
201
N20
>8
P
>99
P
Slight
204
N20
>8
P
>99
P
Slight
SwedishMC Edmonds-DaySurgery-12.8.17-Cut/Cap-Datasht-Gas Outlet Page 1 of 2
Medical Gas Services, LLC
Medical Gas Outlet 1 Inlets Evaluation Worksheet
Facility: Swedish MC - Edmonds Job #: I Date: 12/10/17
Area Tested: Day Surgery I Verifier: HP
Zone Valve - I-1 None n Not Reauired f-1 New N1 Existina
Labeled As: ED West Rooms 17-27
1 ZV Brand: Chemetron
Gas
3 Part Valve
DL Gauge
Ga. PSI
Act PSI
Comments:
OX
❑ Yes ❑ No
❑ Yes ❑ No
54
56
MA
❑ Yes ❑ No
❑ Yes ❑ No
58
55
VAC
❑ Yes ❑ No
❑ Yes ❑ No
20
22
Araa Alarm . I-1 Nnna f-1 Net Reauired I-1 New 1-1 Existina
Labeled As:
I Brand:
Gas
PSI
Set HI
Set LO
Sensor
Comments: Alarm turned off.
OX
MA
VAC
Outlets & Inlets - 1-1 New FRI Existina
Outlet 1 Inlets Brand: Chemetron
Location: Wall
Quick Connect: Chemetron
Area I Room 1 Door
Gas
Flow
SCFM
Delta
Flow
Conc.
%
Part.
Comments:
103
OX
>8
P
>99
P
Slight
MA
>8
P
20.8
P
Slight
VAC
>5
P
--
--
SwedishMC Edmonds-DaySurgery-12.8.17-CuVCap-Datasht-Gas Outlet Page 2 of 2